iChart can only accept imported null data after the initiation fill has been changed to allow that. 1. Open iChart without the current project. 2. From the main menu select Help | Open Users Folder. 3. Go up (1) level in Windows Explorer. 4. Open a file named “ICHART.INI” with notepad. 5. Search for “[Device]”. […]
Articles Tagged: Knowledge Base
Tag new knowledge base for search
Set up a Continuous Alarm in iChart
To have iChart continue to report an alarm while a parameter is outside of a desired range the following steps can be taken. 1. Open iChart with the current project. 2. Click Project | Setup iChart Alarm… from the menu bar. 3. Select the parameter the alarm will be based on from the Parameter list. 4. Click Add. 5. Setup the alarm […]
Separate Date and Time in .csv Export
In some applications the date and time are more useful when separated into two separate columns. 1. Close iChart. 2. Navigate to C:Program FilesNexSensiChart6. 3. Locate the file ICHART.INI. 4. Open ICHART.INI in Notepad. 5. Locate [Report]. 6. Hit <Enter> to create a new line and type the following: SeparaterTimeDataColInCSVFile=1 7. Save the file and close Notepad. 8. […]
Connect a Lufft Weather Sensor for Configuration
In order to make certain configuration changes on a Lufft sensor, it must be connected to the computer directly using a WSConfig cable. 1. Connect the Lufft sensor to the WSConfig cable. 2. Plug the USB cable from the WSConfig cable into the computer. 3. Open Lufft’s Config Tools Software. If not installed, please download […]
Manual Data Manipulation
iChart does not offer a built in data manipulation option. In order to alter the data within iChart one can export the data to Excel, remove the iChart data, and re-import the altered data. 1. Open iChart with the current project file (.icr). 2. Click Help | Open Users Folder from the menu bar. 3. Open the folder named […]
Changing Lufft Weather Sensor Settings in iChart
Changing Lufft local altitude setting via SDI-12 The Lufft line of weather sensors can be configured using SDI-12 protocol through iChart software. 1. Open iChart with the current project. 2. Right click on the Lufft sensor in the Navigation Panel. 3. Select Property. The Smart Sensor Properties window should appear. 4. Record the number in […]
Copying Logger Settings Between Sites
It is possible to copy logger settings – including alarms, alerts, data flags, interrogation schedules and RTC schedules – between sites using iChart software. 1. Open iChart with the current project. 2. In the Navigation Panel, highlight the site from which settings are to be copied by clicking once. 3. Click Project | Copy Logger […]
Setting GPS Bounding Alarms in iChart
In certain monitoring applications, it’s necessary to ensure that equipment stays within defined areas. This is made possible by GPS equipment and bounding alarms that can be set in iChart. 1. Record the upper and lower GPS coordinates of the desired bounding rectangle. GPS1 and GPS2 form two corners of the bounding rectangle in Figure […]
Adjusting Barometric Pressure for Local Elevation
Most barometric pressure sensors measure the absolute pressure of a location and do not account for the elevation above sea level. Since barometric pressure is conventionally reported (in weather reports, etc.) as an adjusted value referenced to sea level elevation, this may cause readings to appear artificially low, particularly at high elevations. To reference sensor […]
Removing the Timestamp in iChart
To reduce the size of each output in order to save time and memory, the timestamp can be removed. 1. Open iChart with a current project. 2. From the main menu select Edit | Preferences. 3. Switch to the Advanced tab. 4. Uncheck the “Send timestamp with each upload cmd” box. Click OK to finish. REV: 13G18