Creating a Generic SDI-12 Sensor Profile on X-Series Loggers
Sensors not available in the current CONNECT library can be deployed on a NexSens X-Series data logger by selecting from a Generic manufacturer and generating a custom profile.
Sensor Setup
Before the sensor can be deployed, complete sensor setup by following the manufacturer’s guidelines.
Note: Be sure to note the parameter order in the software, as this must align with selected parameters in the logger.
Connecting with X-Series Loggers
Next, the sensor will need to be connected to the X-Series logger through CONNECT software. If you have not yet done so, set up the project in WQData LIVE.
- Connect the sensor using the supplied cable to one of the open sensor ports on the logger and note the port number.
- Plug the X3 into the USB port of the computer and power it on with the AC power cable.
- Open CONNECT and select the SENSOR tab.

- Click Add and select Generic from the Manufacturer drop-down. Then select SDI-12 in the model section.
- Adjust the SDI-12 address if needed and select the port number.
- Select either Continuous or Switch in the Power drop-down.
- Add or remove sensor parameters as needed.

- Select Save Sensor Profile.
- Complete the setup for all other sensors, and then click Save.
- Name the Sensor Profile List and select Ok.

- Complete system setup by clicking Deploy.

- Confirm that data is being transferred to WQData LIVE at the next interval (default is 10 minutes).