Creating a Generic SDI-12 Sensor Profile on X-Series Loggers


Sensors not available in the current CONNECT library can be deployed on a NexSens X-Series data logger by selecting from a Generic manufacturer and generating a custom profile.

Sensor Setup

Before the sensor can be deployed, complete sensor setup by following the manufacturer’s guidelines.

Note: Be sure to note the parameter order in the software, as this must align with selected parameters in the logger.

 

Connecting with X-Series Loggers

Next, the sensor will need to be connected to the X-Series logger through CONNECT software. If you have not yet done so, set up the project in WQData LIVE.

  1. Connect the sensor using the supplied cable to one of the open sensor ports on the logger and note the port number.
  2. Plug the X3 into the USB port of the computer and power it on with the AC power cable.
  3. Open CONNECT and select the SENSOR tab.
    Open CONNECT and select the SENSOR tab.
  4. Click Add and select Generic from the Manufacturer drop-down. Then select SDI-12 in the model section.
  5. Adjust the SDI-12 address if needed and select the port number.
  6. Select either Continuous or Switch in the Power drop-down.
  7. Add or remove sensor parameters as needed.
    Generic SGI-12 script generation in CONNECT software.
  8. Select Save Sensor Profile.
  9. Complete the setup for all other sensors, and then click Save.
  10. Name the Sensor Profile List and select Ok.
    Save Sensor Profile List.
  11. Complete system setup by clicking Deploy.
    Generic SGI-12 script generation in CONNECT software.
  12. Confirm that data is being transferred to WQData LIVE at the next interval (default is 10 minutes).