Convert PSI Pressure Output to Level Output in Feet in iChart


The Seametrics PT-12 Submersible Pressure Sensor records the pressure exerted on its internal diaphragm from the surrounding environment and can convert this signal into a level measurement. The X2 data logger script has a built-in conversion factor that outputs the data in both pressure (PSI) and level (ft.) However, for the iSIC and iSIC-V2 data logger, a command must be sent within iChart to input this conversion factor and output level data, in feet, instead of the pressure data in PSI.

 

SDI-12 Commands

  1. Open iChart
  2. Go to Advanced | iSIC | iSIC to connect to the data logger.
Screenshot of iChart iSIC connection interface.

iChart iSIC connection interface (Advanced | iSIC | iSIC).

 

 

  1. Click the SDI-12 tab.
Screenshot of iChart iSIC setup interface with SDI-12 selected.

iSIC Setup SDI-12 tab

 

 

  1. Send ?! to ensure the sensor is connected and can communicate through SDI-12.
    • An acknowledgment (ACK:) should display along with the sensor address.

Screenshot of iChart iSIC setup interface with SDI-12 selected and ?! entered in the COMMAND box.

 

  1. Input the following command to have the pressure output in PSI converted to level in feet, and then click Send.
    • aXC16=2.31!
    • a corresponds to the SDI-12 address of the sensor, which can be found by following the previous step.
Screenshot of iChart iSIC setup interface with SDI-12 selected and aXC16=2.31! entered in the COMMAND box.

SDI-12 level conversion command for PT-12.

 

 

  1. Send the following command to set a level offset of zero.
    • aXC17=0.0!
Screenshot of iChart iSIC setup interface with SDI-12 selected and aXC17=0.0! entered in the COMMAND box.

SDI-12 level offset command for PT-12.

 

 

  1. Add the PT12 as a Generic|SDI-12 sensor in iChart with the attached settings and program the logger.
    1. Note: Match the SDI-12 address of your PT-12 in the smart sensor setup.

 

Using the generic SDI-12 option to add a PT-12 into an iChart project.

Using the generic SDI-12 option to add a PT-12 into an iChart project.