TS210/T-Node FR Sensor Drift


As with all electronics, sensor drift may occur in the NexSens TS210/T-Node FR temperature sensors over time due to aging and exposure to environmental conditions. This drift has been quantified in the laboratory, as shown in the chart below.

Chart depicting the sensor drift of NexSens TS210/T-Node FR thermistors over time.
TS210/T-Node FR Titanium Thermistor.

TS210/T-Node FR Titanium Thermistor.

Temperature is recorded based on the electrical resistance (ohms) through the sensor’s titanium thermistor. As the figure above displays, the resistance changes slightly at the same temperature over time. The “J-shaped” curve suggests that the rate of drift generally decreases with the age of the sensor, but the absolute drift increases with age. Based on the chart, at 50,000 hours (~5.7 years) the nominal drift (assuming 0.014% change) is ~0.003181°C.