The X2 is designed for long-term unattended operation, so minimal maintenance is typically required. The following should be completed periodically depending on equipment use and conditions.

  • Ensure all UW Plug connections are hand tight
    • This ensures a water-tight seal is maintained
  • Ensure antenna connections are tight
    • Water incursion into the antenna connection can lead to low or no signal quality
  • Inspect and grease UW O-Rings
    • Ensure every Plug has an O-Ring around the connector and every Receptacle has an O-Ring at the bottom
    • Inspect O-Rings for cuts or nicks that may allow water to pass
    • Re-grease UW Plug O-Rings every time a connection is made
  • Replace UW O-Rings
    • After several years of use and/or several connection¬†cycles, it is recommended to replace UW O-Rings to ensure a water-tight seal is maintained
  • Clean Unit
    • Using a damp cloth, wipe any dirt or debris from the unit