Sea-Bird Scientific HydroCAT CTD X2 Integration Guide


  1. Follow the wiring diagram below to connect your Sea-Bird Scientific HydroCAT CTD to a PC by utilizing the Sea-Bird manufacturer included MCIL-6-FS to DB9 software cable. An RS232-to-USB converter will be necessary to connect to a PC.
    1. Warning: Ensure the red wire (voltage in) is only wired for the sensor, and has no wiring connection with the RS-232-to-USB converter.
MCIL-6-FS Pin #DB9 Serial Pin #External Wire ColorSignal
15Black + Spare (Both black external wires)GND
23RS232 Rx
32RS232 Tx
4YellowSDI-12
5Reserved
6RedVoltage In

 

  1. Follow the link provided to download the Sea-Bird Scientific UCI software for MAC or Windows.
    1. Open the software once downloaded.

 

UCI software icon

 

  1. Under the UCI Dashboard tab, click Connect. A separate Connect window will open. Enter the following settings:
    1. Instrument Type: HydroCAT
    2. Baud Rate: 19200
    3. Correct COM port

 

  1. Click Connect when the settings are input.

Click connect to begin communicating with the sensor.

 

  1. Once connected, the Connection Mode at the top left will display Setup in green, along with the internal sensor diagnostics.

Successful sensor connection is indicated by the green ‘Setup’ indicator.

 

  1. Open the HydroCAT command terminal by going to Sensor | HydroCAT (no optics) | Command Terminal. 

Access the HydroCAT command terminal.

HydroCAT command terminal with the ‘GetCD’ command input.

 

  1. Set the output format to SDI-12 Engineering by selecting OutputFormat=[value] from the drop-down menu. Replace [value] with the number 3 and click Submit.
    1. The command terminal should successfully display <Executed/> before and after the command.

Successful SDI-12 Engineering output command.

 

  1. Go to the HydroCAT Settings by following HydroCAT (no optics) | HydroCAT Settings.

Open the HydroCAT Settings.

 

  1. Modify the output format to the following parameters and units:
ParameterUnits
Temperature°C
ConductivityS/m
PressureDecibar
Oxygenmg/L
Salinitypsu
Sound Velocitym/sec
Specific ConductivityS/m

 

Correct output format.

 

  1. Return to the Command Terminal and enter ts into the command line and  press Submit.
    1. This command should return the following output in order, separated by either a + or – sign (depending on a positive or negative measurement):
      • SDI-12 Address
      • Temperture
      • Conductivity
      • Pressure
      • Oxygen
      • Salinity
      • Sound Velocity
      • Specific Conductivity
        • e.g. SDI-12 Address + Temperature – Conductivity – Pressure + Oxygen + Salinity + Sound Velocity – Specific Conductivity

ts output command.

 

  1. Once the sensor output has been changed, a Generic SDI-12 script will need to be generated in the CONNECT software following the outline below:
    1. Manufacturer: Sea-Bird
    2. Model: HydroCAT SDI12
    3. SDI-12 Address: Must match the address input in the UCI software.
    4. Port: Must match the port in which the HydroCAT will be detected.
    5. Warmup: Recommended 30 seconds.
    6. Meas Cmd: Set as the ‘C’ command to avoid any interference with other SDI-12 sensors.
    7. Power: Recommended switch power
    8. Selected Parameters: Must match the parameter configuration set in the UCI software.

Sea-Bird HydroCAT generic SDI-12 CONNECT script.

 

  1. Once completed, transfer the script file onto the X2 and follow the detection process outlined in your included X2 Quick Start Guide, unless the X2 detection has been completed by NexSens prior to shipment.