Click to download MSDS for WQ-BR Smart USB Sensor REV: 13J18
WQ-BR Bromide Smart USB Sensor
WQSensors Troubleshooting Tips
Follow procedures below to isolate and resolve sensor problems. Symptom Possible Cause Corrective Action Out of range calibration Considerable temperature gradient between standard solutions and reference fill solutions Soak the electrode in the low range calibration standard for two hours and recalibrate Contamination of calibration standards, poor dilution accuracy or distilled water quality, mathematical error […]
WQ-BR Bromide Smart USB Sensor Maintenance
Maintenance If the response time and/or quality of data collected by a WQSensor decrease, try cleaning and reconditioning the probe. If these steps do not resolve the problem, consider replacing the sensing module to boost performance. Cleaning The sensor can be cleaned using distilled or deionized water. Rinse thoroughly as necessary but avoid using any […]
WQSensors Built-in Customer Support
WQSensors software integrates the internet into part of its functionality. Software updates, a technical support interface, WQSensor user manuals, and a WQSensor shopping interface can all be accessed through the Help tab. If phone contact is preferable, include a note in the Message field of WQSensors Software Help suggesting a call back. Be sure […]
WQSensors Science Library
The WQSensor software Library menu is a desktop reference for water, wastewater, and environmental professionals. It’s packed with tables, calculations, and many conversions. Among the science library’s capabilities are the following: Converting units of measure Calculating flow over a weir Determining the inside diameter of PVC pipe Finding the ammonia tolerance level for rainbow trout […]
WQSensors Optional Software Settings
Click the Application button in the Settings tab to toggle the Application Settings described below on or off as desired. “Show introduction animation?” – displays an introductory animation each time WQSensors software is started. Un-checking this box will disable animation. “Remember last menu location?” – specifies whether the WQSensors software should reload the […]
WQSensors Project Files
Each project file contains information about every sensor that was connected to the computer while that file was open, as well as any data that was collected during that time. When WQSensors software first runs, it defaults to “Project_1”. This is the default name for the first project file. To create a new […]
Good Laboratory Practice Files
WQSensors software automatically records every calibration to a Good Laboratory Practice (GLP) file. The GLP file includes date & time, calibration values, and mV values for each calibration. This file is stored on the sensor and will carry over between computers. The GLP files can be accessed for any sensor connected to the computer by […]
Calibrating WQ-BR Bromide Smart USB Sensor
It is recommended that calibration is done on a daily basis, or at least prior to each use. During calibration, observe the slope of the mV readings between the two points that are being used for calibration. Slope is defined as the change in voltage read by the sensor with a ten-fold change in concentration […]
WQSensors Reporting & Exporting Data
WQSensors software provides an interface to view data in raw or graphical form, automatically summarize data, compute statistical parameters, and export data. Any data that is logged can be viewed or exported by generating a report. Note that data is exported in .csv format for use with programs like Microsoft Excel. To generate a report, […]